Radical-7 Signal Extraction Pulse CO-Oximeter Operator’s Manual 7-5
7
SERIAL INTERFACE SETUP (CONTINUED)
The pin-outs for the RS-232 connector are shown in the following table:
PIN SIGNAL NAME
1 No Connection
2 Receive data – RS-232 ±9 V (±5 Vmin)
3 Transmit data – RS-232 ±9 V (±5 Vmin)
4 No Connection
5 Signal Ground Reference for COM signals
6 No Connection
7 No Connection
8 No Connection
9 No Connection
SERIAL PRINTER SETUP
To print the SpO
2
and pulse rate data in ASCII 1 format on a serial printer, simply connect the
serial printer to the serial port. Once serial communication is established, the Radical-7 Pulse
CO-Oximeter will automatically start printing the ASCII 1 text data.
WARNING: ALL EXTERNAL INSTRUMENT CONNECTIONS TO THE RS-232
SERIAL PORT MUST BE IEC-60950 COMPLIANT.
Analog Output/Nurse Call Specifications
The Analog Out and Nurse Call are features accessible on the same female high density
DB-15 connector.
NOTE: The Radical-7 Pulse CO-Oximeter analog output / nurse call interface is only available when
the Radical-7 Pulse CO-Oximeter Handheld is properly attached to the Radical-7 Pulse CO-Oximeter
Docking Station. Only use an analog / nurse call cable that has a ferrite bead installed.
NOTE: The analog output / nurse call interface is not available in all versions of the Docking Station.
The following table shows the pinout of the analog output and nurse call.
PIN SIGNAL NAME
1
+5V (60mA max.)
2
Ground
3
Ground
4
Ground
5
Ground
6
Nurse Call (Normally Open)