7-4 Radical-7 Signal Extraction Pulse CO-Oximeter Operator’s Manual
7
4 The Radical-7 has been validated for low perfusion accuracy in bench-top testing against a Biotek Index 2
TM
* simulator and Masimo's
simulator with signal strengths of greater than 0.02% and transmission of greater than 5% for saturations ranging from 70-100%. This
variation equals plus or minus one standard deviation. Plus or minus one standard deviation encompasses 68% of the population.
5 Masimo Rainbow SET technology with Masimo sensors has been validated for pulse rate accuracy for the range of 25-240 bpm in
bench top testing against a Biotek Index 2 simulator. This variation equals plus or minus one standard deviation which encompasses
68% of the population.
6 SpHb accuracy has been validated on healthy adult male and female volunteers and on surgical patients with light to dark skin
pigmentation in the range of 8 g/dL to 17 g/dL SpHb against a laboratory CO-Oximeter. The variation equals plus or minus one standard
deviation which encompasses 68% of the population. The SpHb accuracy has not been validated with motion or low perfusion.
7 This represents approximate run time at the lowest indicator brightness and pulse tone turned off using a fully charged battery.
8 If the batteries are to be stored for extended periods of time, it is recommended that they be stored between -20°C to +30°C, and at a
relative humidity less than 85%. If stored for a prolonged period at environmental conditions beyond these limits, overall battery capacity
may be diminished, and lifetime of the batteries may be shortened.
9 With FastSat the averaging time is dependent on the input signal. For the 2 and 4 second settings the averaging time may range from
2-4 and 4-6 seconds, respectively.
10 Maximum sensitivity mode fixes perfusion limit to 0.02%.
*Registered trademark of Fluke Biomedical Corporation, Everett, Washington.
Serial Interface Specifications
The digital interface for serial communication is based on the standard RS-232 protocol.
The Radical-7 Pulse CO-Oximeter by default always outputs ASCII 1 text data through the
serial port, unless the user selects a different output mode in the Output menu. To interface
with the Radical-7 Pulse CO-Oximeter and receive serial text data, simply connect a serial
interface cable with a ferrite bead installed to the serial output connector located on the back
of the Radical-7 Docking Station.
NOTE:
The Radical-7 Pulse CO-Oximeter serial interface is only available when the Radical-7 Pulse
CO-Oximeter Handheld is properly attached to the Radical-7 Pulse CO-Oximeter Docking Station.
NOTE: The serial interface is not available in all versions of the docking station.
Once serial communication is established, packets of data are communicated at 1 second
intervals. The data packets contain: the date, time, SpO
2
, SpMet, SpCO, SpHb, SpOC, pulse
rate, perfusion index, PVI, and alarm and exception values (in ASCII format).
WARNING:
ALL EXTERNAL INSTRUMENT CONNECTIONS TO THE ANALOG OUTPUT/
NURSE CALL CONNECTOR MUST BE IEC-60950 COMPLIANT.
SERIAL INTERFACE SETUP
To interface with the Radical-7 Pulse CO-Oximeter serial port, set the following
communication parameters on the interfacing serial instrument:
PARAMETER SETTING
BAUD RATE
9600 Baud bi-directional
NUMBER OF BITS PER CHARACTER
8
PARITY
None
BITS
1 start, 1 stop
HANDSHAKING
None
CONNECTOR TYPE
Female DB-9
Specifications
Radical-7 Specifications (continued)