Radical-7 Signal Extraction Pulse CO-Oximeter Operator’s Manual 2-5
2
System Description
HANDHELD FRONT PANEL (CONTINUED)
22
13
SpOC*
MEASUREMENT
DISPLAY
The measurement of total oxygen content is displayed in units of
milliliter per deciliter.
23
SENSOR 4 Hrs
SENSOR TIME
Sensor Time displays the amount of time remaining (for SpHb
Rainbow Reusable sensors) in the lower, right-hand corner of the
screen. The Sensor Time is displayed in “X” Hrs” or “XX Min.” As the
sensor monitors the patient, the amount of time decreases to “0”.
Once the sensor time shows ”0”, the sensor must be replaced.
The Sensor Time and the Time and Date Indicator show in the
same location when the Time and Date Indicator is set to show on
the screen. Subsequently, the Sensor Time replaces the Time and
Date Indicator for 2 minutes at the following Sensor Times: 4 hours,
2 hours, 1 hour and 0 minutes. Refer to Section 4, Sensor Time
Remaining for additional information. Once the Sensor Time reads 2
minutes, the Sensor Time will continue showing as it decreases to “0”.
10/11/06
12:23:11am
TIME AND DATE
INDICATOR
The Time and Date Indicator displays the current time and date. The
time is displayed in 12 or 24 hour format. The date is displayed in dd/
mm/yy or mm/dd/yy format. Select the date and time display formats
in the Clock menu.
24
5
PLETH
VARIABILITY
INDEX
PVI is displayed as a percentage. The lower the number, the less vari-
ability there is in the PI over a respiratory cycle.
25
SIGNAL IQ
The Signal IQ shows the acquired signal quality and the timing of
the pulse. A tall vertical line indicates a high quality signal, while a
short vertical line indicates a low quality signal. The Signal IQ will
be displayed as a single, pulsating bar in the Numbers mode.
26
PULSE WAVEFORM
DISPLAY
The Pulse Waveform Display shows the acquired plethysmographic
waveform. The plethysmographic waveform is scaled with signal
strength. Signal strength is defined as the relation of arterial pulsatile
signal to the non-pulsatile signal component.
27
BRIGHTNESS
LEVEL
The Brightness level icon displays when utilizing the Backlight/
Contrast button.
28
3.00
PERFUSION INDEX
The Perfusion Index indicates numerically the percentage of
pulsatile signal to non-pulsatile signal (pulse strength).
29
APOD
SENSITIVITY
The sensitivity icon is shown on the Radical-7 display to indicate
if the Radical-7 is set to operate in Normal, Maximum (MAX) or
Adaptive Probe Off Detection (APOD) mode. When in Normal mode,
this area will appear blank.
30
140
50
PULSE RATE
ALARM LIMITS
DISPLAY
The Pulse Rate Alarm Limits Display shows the upper and lower
pulse rate alarm limits. When an alarm limit is exceeded, the pulse
rate value and the violated limit flashes.
31
FastSat
FASTSAT
The FastSat
®
label is shown on the Radical-7 display whenever the
Radical-7 is set to operate in the FastSat mode.
32
76
PULSE RATE
The Pulse Rate Measurement Display shows the patient’s pulse rate
in beats per minute. The upper and lower pulse rate alarm limits are
also displayed next to the pulse rate measurement. The pulse rate is
calculated and the display is updated at a frequency of once per second.
* For instruments that include SpMet, SpCO and SpHb parameters: the SpMet, SpCO, SpHb and SpOC parameters/
measurements will be dimly lit in the display screen if a patient cable is not attached or a non-Rainbow sensor is being used.
NOTE: The Pulse CO-Oximeter display view will be different without the optional SpCO, SpMet and SpHb
parameters/measurements.