Radical-7 Signal Extraction Pulse CO-Oximeter Operator’s Manual 3-7
3
THE RADICAL-7 REVERTS TO INTERFACE ALARM SpO
2
/BPM “Yes” DURING
POWER INTERRUPTIONS OR WHEN THE SATSHARE CONNECTION IS LOST
OR THE Instrument BECOMES SEPARATED FROM THE DOCKING STATION.
THIS ENSURES THAT THE RADICAL-7 PROVIDES AUDIBLE ALARMS FOR SpO
2
AND PULSE RATE WHEN THE CONNECTION TO THE INTERFACED SYSTEM
BECOMES COMPROMISED.
Philips VueLink Setup
CAUTIONS:
 SIMULTANEOUS USE OF SATSHARE AND SERIAL PORT IS NOT SUPPORTED.
 IF THE RADICAL DOCKING STATION IS COMPATIBLE WITH PATIENT SAFETYNET
OR RADNET
TM
, VUELINK IS NOT SUPPORTED.
 ONLY SpO
2
AND PULSE RATE CAN BE DISPLAYED ON THE MULTIPARAMETER
MONITOR WITH VUELINK.
1. Select the Philips VueLink selection from the Output menu on the Radical-7 Pulse
CO-Oximeter. Refer to Section 4, Output.
2. Connect one end of the VueLink cable to the Serial Output connector on the back of
the Docking Station.
3. Connect the other end of the VueLink cable to the VueLink module and insert the
module into the Philips monitor rack.
4. The SpO
2
and pulse rate values will automatically appear on the Philips monitor.
5. In order for the plethysmographic waveform to be displayed on the Philips monitor
and for the Philips monitor to indicate the alarm conditions measured by the Pulse
CO-Oximeter, the user must configure the Philips monitor. Refer to the Philips
Operator’s Manual for complete instructions.
6. The Radical-7 can be set up to audibly indicate all patient alarms while communicating
with the Philips VueLink module. Use the Interface Alarms setting in the General menu
to enable and disable audible alarms on the Radical-7.
Spacelabs Universal Flexport Setup
CAUTIONS:
 SIMULTANEOUS USE OF SATSHARE AND SERIAL PORT IS NOT SUPPORTED.
 ONLY SpO
2
AND PULSE RATE CAN BE DISPLAYED ON THE MULTIPARAMETER
MONITOR WITH FLEXPORT.
1. Select the Spacelabs Flexport selection from the Output menu on the Radical-7.
2. Connect one end of the Spacelabs Flexport cable to the Serial Output connector on
the back of the Docking Station.
3. Connect the other end of the Spacelabs Flexport cable to the Spacelabs Universal
Flexport connector.
4. The SpO
2
and pulse rate values will automatically appear on the Spacelabs screen.
WARNING: EXTERNAL INSTRUMENT CONNECTIONS TO THE SATSHARE
PORT MUST BE IEC-60601-1-1 COMPLIANT.
Setup