4-26 Radical-7 Signal Extraction Pulse CO-Oximeter Operator’s Manual
4
Operation
Trend Display
Once the Trend Display touch key icon is selected, the trend data is displayed on the main screen.
The Radical-7 Pulse CO-Oximeter stores one data set of SpO
2
, pulse rate, SpMet, SpCO,
SpHb, SpOC, PI, PVI and system messages in a dedicated memory area. Depending on the
Trend Period, a setting for how often the data is stored in the trend memory, the Radical-7 Pulse
CO-Oximeter can store between 72 hours and > 10 days of trend data. The Radical-7 Pulse
CO-Oximeter also employs a sophisticated data compression scheme. The actual amount of
trend data that is stored is dependent on the type of data that is collected.
The Radical-7 Pulse CO-Oximeter only stores data in the trend memory while the instrument is
turned on, and the trend data remains in memory until the memory is full, or is cleared by the user.
CAUTION: CHANGING THE DATE AND TIME OF THE SYSTEM CLOCK, OR CHANGING THE
TREND PERIOD, WILL ALSO CLEAR THE DATA IN THE TREND MEMORY.
The following table outlines the trend capacity for sample Trend Period settings:
TREND PERIOD TREND MEMORY CAPACITY
2 SECONDS
UP TO 72 HOURS (APPROXIMATELY 3 DAYS)
10 SECONDS
> 10 DAYS
The Trend Display can be configured to display one or two of any of the available trend
parameters/measurements (SpO
2
, SpMet, SpCO, SpHb, SpOC, pulse rate, PI or PVI) selected
by the user. The instrument is storing all parameters/measurements in trend memory, but can
only display one or two user selected parameters/measurements at any one time. The Trend
Display can be adjusted to the desired parameter/measurement by selecting Trend from the
Main, Display Menu.
TREND AUTO SCALE
The Radical-7 can be set to automatically change the trend data field by adjusting the area
(scale) of the graph that is shown. The scale of the trended graph is adjusted to focus on the
values of the data presented. The movable area of focus allows improved data viewing. As
the data trends up or down, the Radical-7 automatically adjusts the scale (or viewing area) to
maintain the data within the field of view.
TREND DISPLAY SCREEN