Radical-7 Signal Extraction Pulse CO-Oximeter Operator’s Manual 3-5
3
SatShare Setup
The Radical-7 Pulse CO-Oximeter has been proven to be accurate during patient motion and low
perfusion conditions. Saturation and pulse rate values from the Radical-7 Pulse CO-Oximeter may
be displayed on a multiparameter monitor through the SatShare feature.
The SatShare feature provides an ideal, simulated waveform corresponding to the measured
saturation and pulse rate values determined by the Masimo SET technology. This waveform
may be used to display these values on multiparameter monitors through the multiparameter
oximetry sensor or input connector.
It is recommended that the Radical-7 Pulse CO-Oximeter is positioned close to the
multiparameter monitor with the Radical-7 Pulse CO-Oximeter screen visibly displaying the
plethysmographic waveform and the saturation and pulse rate measurements.
CAUTION:
SIMULTANEOUS USE OF SATSHARE AND SERIAL PORT IS NOT SUPPORTED.
97
76
SpHb g/dl
17
7
10.0
SpO
C
SATSHARE SETUP
1. Select the SatShare cable that is appropriate for the multiparameter monitor that is
being connected. Check the Masimo web site at www.masimo.com for the latest list of
available SatShare cables and validated instruments.
2. Connect the labeled end of the cable to the SatShare Cable Connector port on the
back of the Docking Station. Tighten the connector screws for a secure connection.
3. Connect the other end of the SatShare cable either to the sensor connector of the
multiparameter monitor’s SpO
2
cable or directly to the SpO
2
connector on the monitor.
4.
Verify that the Radical-7 Pulse CO-Oximeter recognizes the correct cable. The name of the
SatShare cable will be displayed on the LCD screen when the SatShare mode is functional.
5. Set the multiparameter monitor’s high and low saturation and pulse rate alarm limits
as appropriate.
6.
Set the multiparameter monitor’s averaging time to the lowest setting (i.e. fastest response). The
Radical-7 Pulse CO-Oximeter’s ideal waveform necessitates the need for additional averaging
by the monitor. If the multiparameter monitor’s averaging time is not changed, the time to display
physiological changes in saturation on the monitor will be increased with SatShare. However, the
delay can be minimized by reducing the multiparameter monitor’s averaging time.