3-6 Radical-7 Signal Extraction Pulse CO-Oximeter Operator’s Manual
3
SATSHARE SETUP (CONTINUED)
7. While in the SatShare mode, if there are any significant discrepancies between the
readings from the Radical-7 Pulse CO-Oximeter and those on the monitor displaying
the values obtained from SatShare, the values reported by the Radical-7 Pulse
CO-Oximeter are to be considered the correct values.
8. To use the Radical-7 Pulse CO-Oximeter with SatShare while it is not connected to
AC power, set the Power Save parameter in the General menu to “No” and refer to
Section 4, Operation. Please note that if the Radical-7 Pulse CO-Oximeter is used in
this mode, the length of time the Radical-7 Pulse CO-Oximeter can operate on battery
power will be significantly diminished.
9. Set the SatShare Numbers and the Interface Alarms SpO
2
/pulse rate parameters in
the General menu according to customer preference.
When Rainbow parameters (SpCO, SpMet, SpHb, etc.) are configured, Interface
Alarms "SpO
2
/BPM" can be set to “Yes” or “No”. The “Yes” setting allows SpO
2
and
pulse rate audible alarms at both the Radical-7 and the interfaced system. The “No”
setting mutes the SpO
2
and pulse rate audible alarms at the Radical-7 while allowing
SpO
2
and pulse rate audible alarm alerts at the interfaced system. The “No” setting
prevents both systems (Radical-7 and interfaced system) from producing audible
alarms at the same time. See section 4: General.
NOTE: The Radical-7 reverts to Interface Alarm "SpO
2
/BPM" “Yes” during power
interruptions or when the SatShare connection is lost or the instrument becomes
separated from the docking station. This ensures that the Radical-7 provides audible
alarms for SpO
2
and pulse rate when the connection to the interfaced system
becomes compromised.
10. If displaying the simulated waveform is not desirable, it is recommended to turn off the
plethysmographic waveform display of the multiparameter patient monitor.
SATSHARE SIGNALS ARE IDEAL SIMULATED WAVEFORMS CORRESPONDING TO
THE CALCULATED SATURATION AND PULSE RATE VALUES AND DO NOT CONTAIN
ALL OF THE INFORMATION CONTAINED IN PHYSIOLOGICAL WAVEFORMS. THE
MULTIPARAMETER PATIENT MONITOR TRANSLATES THESE SIGNALS INTO
SATURATION AND PULSE RATE VALUES.
DURING SATSHARE OPERATION, THE AUDIBLE ALARMS MAY BE MUTED ON
THE RADICAL-7 PULSE CO-OXIMETER. WHEN THE AUDIBLE ALARM IS MUTED
(INDICATED BY THE BELL WITH A SLASH THROUGH IT) ON THE RADICAL-7
PULSE CO-OXIMETER, USE THE MULTIPARAMETER MONITOR FOR AUDIBLE
ALARM INDICATION.
DURING SATSHARE OPERATION DO NOT USE THE PLETHYSMOGRAPHIC
WAVEFORM DISPLAY ON THE MULTIPARAMETER MONITOR FOR DIAGNOSTIC
PURPOSES. INSTEAD, USE THE PLETHYSMOGRAPHIC WAVEFORM DISPLAYED
ON THE RADICAL-7 PULSE CO-OXIMETER SCREEN.
TO AVOID EXCESSIVE BATTERY DISCHARGING, DO NOT CONNECT ANY
EQUIPMENT TO THE SATSHARE CONNECTOR UNLESS THE RADICAL-7 PULSE
CO-OXIMETER IS CONNECTED TO THE AC MAINS POWER SUPPLY.
ONLY USE A SATSHARE CABLE THAT HAS A FERRITE BEAD INSTALLED.
ONLY SpO
2
AND PULSE RATE CAN BE DISPLAYED ON THE MULTIPARAMETER
MONITOR WITH SATSHARE.