1-8 Radical-7 Signal Extraction Pulse CO-Oximeter Operator’s Manual
1
MASIMO SET DST
®
SpOC
12
FST
MST
Overview
SpCO, SpMet AND SpHb
MEASUREMENTS DURING
PATIENT
MOTION
The Radical-7 displays measurements of SpCO, SpMet and SpHb during patient motion. However,
because of the changes in the physiological parameters such as blood volume, arterial-venous
coupling, etc. that occur during patient motion, the accuracy of such measurements may not be
reliable during excessive motion. The measurements for SpCO, SpMet and SpHb display “---” and a
message, “LOW SpCO SIQ”, “LOW SpMet SIQ” or “LOW SpHb SIQ”, displays to alert the clinician
that the instrument does not have confidence in the value due to poor signal quality caused by
excessive motion or other signal interference.
MASIMO RAINBOW SET PARALLEL ENGINES
This fi gure is for conceptual purposes only.