58 Radical Signal Extraction Pulse Oximeter Operator’s Manual
9
service/maintenance
battery operation and maintenance
The Radical Handheld includes a 1.5 Amp-Hour Nickel Metal Hydride battery. The Radical
Docking Station may include the optional 6.5 Amp-Hour Nickel Metal Hydride battery.
Before using the Radical as a Handheld or transport monitor, the Handheld battery and the
optional Docking Station battery need to be fully charged.
To charge the battery(s), attach the Handheld unit to the Docking Station. Ensure that AC
power is attached to the Docking Station.Turn the AC power switch to the l position.Verify
that the battery(s) are charging; the battery charging LED indicators on the Docking Station
flash prior to charging and remain illuminated while the battery(s) are charging. A
continuously flashing battery charging LED indicates that the internal battery temperature
exceeds recommended operating conditions for proper battery charging. Proper battery
charging will proceed when the temperature returns to recommended operating conditions.
The Handheld battery requires approximately 2 to 3 hours for charging. The optional
Docking Station battery requires approximately 6 hours for charging.
When the battery charging LED indicators turn off, additional trickle charging may occur to
complete charging. Although battery charging can occur while the Handheld is docked and
powered on, most efficient charge times are achieved with the Handheld unit turned off.
CAUTIONS:
ALL BATTERIES LOSE CAPACITY WITH AGE, THUS THE AMOUNT OF RUN
TIME LEFT AT LOW BATTERY WILL VARY DEPENDING UPON THE AGE OF
THE BATTERY.
AT LOW BATTERY CONNECT THE RADICAL TO AC POWER TO PREVENT
LOSS OF POWER.
During battery operation of the Radical, please note that the following operating conditions
affect the estimated run-time of the included batteries:
Illumination of the backlit LCD screen. To conserve battery power, keep the backlit
LCD screen at minimum illumination.
Volume of the alarm tones. To conserve battery power, keep the frequency of the
audible alarms to a minimum and at minimum volume.
The SatShare feature.To conserve battery power, always keep the unit on AC line power.
Memory effects of the batteries may shorten run-time of the batteries. As a precaution
against memory effects, it is advisable to completely discharge and fully recharge the
batteries at least once every six (6) months. To properly discharge the batteries, use the
Battery Discharge function, as described in Section 4, Service.