Radical Signal Extraction Pulse Oximeter Operator’s Manual 21
4
introduction
To operate the Masimo SET Radical Pulse Oximeter effectively, the device must be set up
properly, and the operator must:
Know how the pulse oximeter derives its readings (see Section 1, Pulse Oximetry).
Be familiar with its controls, components and operation.
Understand its status and alarm messages (see Section 5, Alarm Identification,
System Messages and Section 6, Troubleshooting).
basic operation
GENERAL SETUP AND USE
1. Inspect the oximeter case for damage.
2. Connect the patient cable to the Patient cable connector of the Radical. Make sure
it is a firm connection and the cable is not twisted, sliced or frayed.
3. Ensure that the power cord is plugged into the Power Entry Module of the oximeter
and into the AC power.
4. Ensure that the AC power switch of the Power Entry Module of the Docking Station
is in the | position.
NOTE: Some power entry modules may not leave the on/off switch.
5. Select a sensor that is compatible with the oximeter before connecting it to the
patient cable. See section 8, Sensors and Patient Cables. If using a single patient
adhesive or disposable sensor, check that the emitter (red light) and the
photodetector (white plastic casing) are properly aligned. If using a reusable
sensor, make sure it opens and closes smoothly. Remove any substances that
may interfere with the transmission of light between the sensors light source and
photodetector.
6. Attach the sensor to the patient. Refer to the Directions for Use of the sensor.
7. Connect the sensor to the patient cable with the logos lining up; make sure it is a
firm connection.
8. Press the Power/Standby button to turn the oximeter on.
9. Make sure the display window is free of alarm and system failure messages (see
Section 5, Alarm Identification).
10. On the display, verify
The high and low alarm limits for SpO
2
and pulse rate.
The readings for SpO
2
and pulse rate.
NOTE: “—” may appear on the numeric display until the SpO
2
and pulse rate
readings have stabilized (approximately 10 seconds).
operation