Radical Signal Extraction Pulse Oximeter Operator’s Manual 41
4
■
The user can access the menu system.
■
If the SatShare cable is connected to the Radical only, and not to a patient monitor,
the SatShare cable type is flashing on the LCD screen.
■
Once Radical detects the presence of a patient monitor, the SatShare cable type
remains constantly displayed on the LCD screen.
■
Patient Alarms of the multiparameter patient monitor will be triggered by the alarm
setting of the patient monitor and not the Radical. To synchronize the alarm events
set the alarm limits of the Radical to those of the patient monitor, or vice versa.
■
Once Radical detects that the SatShare cable is disconnected from the patient
monitor, or if the patient monitor is turned off, Radical automatically returns to normal,
standalone operation.
■
In the SatShare mode, the pulse beep tone of the Radical is disabled.
■
The Radical may automatically set the averaging time during SatShare operation.
For averaging times of 10 seconds and higher, the Radical will automatically set the
averaging time to 8 seconds during SatShare operation. Averaging times of 2, 4 or
8 seconds remain unchanged during SatShare operation. When the Radical returns
to non-SatShare operation, the Radical will maintain the averaging time setting
used in the SatShare mode.
CAUTIONS:
■
SATSHARE SIGNALS ARE IDEAL SIMULATED WAVEFORMS
CORRESPONDING TO THE CALCULATED SATURATION AND PULSE RATE
VALUES AND DO NOT CONTAIN ALL OF THE INFORMATION CONTAINED IN
PHYSIOLOGICAL WAVEFORMS. THE MULTIPARAMETER PATIENT MONITOR
DECODES THESE SIGNALS INTO SATURATION AND PULSE RATE VALUES.
■
DURING SATSHARE OPERATION, THE AUDIBLE PATIENT ALARMS MAY BE
AUTOMATICALLY DISABLED AND SILENCED. USE THE MULTIPARAMETER
PATIENT MONITOR ALARMS TO INDICATE ALARM CONDITIONS.
■
DURING SATSHARE OPERATION DO NOT USE THE PLETH WAVEFORM
DISPLAY ON THE MULTIPARAMETER MONITOR FOR DIAGNOSTIC
PURPOSES. INSTEAD, USE THE PLETH WAVEFORM DISPLAYED ON THE
RADICAL SCREEN.
To return from SatShare operation to normal standalone operation, simply disconnect the
SatShare cable from the patient monitor or disconnect the SatShare cable from the
SatShare connector on the back of the Radical.