Radical Signal Extraction Pulse Oximeter Operator’s Manual 45
5
alarms/messages
system messages
The following chart alphabetically lists all system messages displayed on the LCD
screen. The cause of the message, and the action(s) to be taken are also shown.
The operator should become thoroughly familiar with this information before using the
oximeter for patient monitoring.
MESSAGE POSSIBLE CAUSE(S) RECOMMENDATION
AMBIENT LIGHT Too much light on Remove or reduce lighting.
patient (sensor). Cover sensor from light.
Inadequate tissue Reposition sensor.
covering sensor detector.
DEFECTIVE Oximeter cannot identify Broken sensor cable wire or
SENSOR the connected sensor inoperative LEDs or faulty detector;
the sensor has failed. Replace sensor. Refer to the instructions for
the sensor being used.
INTERFERENCE Outside signal or energy Remove outside interference.
preventing reading.
INVALID SENSOR Oximeter cannot identify Broken sensor cable wire or inoperative
the connected sensor LEDs or faulty detector; the sensor has
failed. Replace sensor. Refer to the
instructions for the sensor being used.
LOW BATTERY Battery charge is low. Charge battery by placing the Radical
Handheld into the Docking Station and
powering the unit with AC line power.
Replace battery if necessary.
LOW PERFUSION Signal too small. Move sensor to better perfused site. Refer
to Section 4, Low Perfusion.
LOW SIGNAL IQ Low signal quality. Ensure proper sensor application. Move
sensor to a better perfused site. Refer to
Section 4, Signal IQ.