Rad-57 Signal Extraction CO-Pulse Oximeter Operator’s Manual 7-1
7
specifications
Rad-57 family specifications
PERFORMANCE
measurement range
Oxygen Saturation (%SpO
2)
: 1-100%
Carboxyhemoglobin Saturation (%SpCO ) 1 - 99%
Pulse Rate (bpm): 25-240 beats per minute
Perfusion Index: 0.02 - 20%
ACCURACY
Oxygen Saturation Accuracy (%SpO
2
) during no motion conditions
1
Adults, Pediatrics 70% - 100% ±2 digits
1% - 69% unspecified
Neonate 70% - 100% ±3 digits
1% - 69% unspecified
Oxygen Saturation Accuracy (%SpO
2
) during motion conditions
2
Adults, Pediatrics
2
70% - 100% ±2 digits
1% - 69% unspecified
Neonate 70% - 100% ±3 digits
1% - 69% unspecified
Carboxyhemoglobin Saturation Accuracy (%SpCO)
3,7
1% - 40% ±3 digits
Pulse Rate (bpm) during no motion conditions
1
Adults, Pediatrics, Neonate 25 to 240 ±3 digits
Pulse Rate (bpm) during motion conditions
2
Adults, Pediatrics, Neonate 25 to 240 ±5 digits
Low Perfusion Performance
3
>0.02% Pulse Amplitude and Oxygen Saturation
(%SpO
2
) ±2 digits
% transmission > 5% Pulse Rate ±3 digits
Resolution
Oxygen Saturation (%SpO
2
)1%
Carboxyhemoglobin saturation (%SpCO), digital display 1%
Carboxyhemoglobin saturation(%SpCO), continuous bar display 5%
Pulse Rate (bpm) 1 bpm
Interfering Substances
Carboxyhemoglobin may erroneously increase oxygen saturation readings. The level of increase
is approximately equal to the amount of carboxyhemoglobin present. Dyes, or any substance
containing dyes, that change usual arterial pigmentation may cause erroneous readings.