4-8 Radical-7 Color Display Signal Extraction Pulse Oximeter Operator’s Manual Radical-7 Color Display Signal Extraction Pulse Oximeter Operator’s Manual 4-94-9
4
operation
Navigating the Main Menu
When the main menu is accessed, the plethysmograph and Signal IQ waveform displays
are replaced with the main menu items. The touch key icons, displayed along the right edge
of the LCD display, are also replaced by the menu access icons. When the main menu is
accessed the monitor remains functional and the saturation and pulse rate numbers will
continue to be displayed.
MAIN MENU SELECTION
The top menu category uses the following four menu selections and touch key control
buttons and icons.
EXIT
Select the Exit icon to exit the main menu.
SELECT
CATEGORY
Select the Select Category icon to select the highlighted menu
item and enter the next level menu.
PREVIOUS
Select the Previous icon to scroll through the menu items without
selecting them. Once a menu item is highlighted, enter the menu
by pressing the Select Category icon.
NEXT
Select the Next icon to scroll through the parameters. Once a
parameter is highlighted, edit the parameter by pressing the Edit
Parameter icon.
MENU CATEGORIES
Once a menu category has been selected, a new set of menu selections and icons are
displayed.
EXIT
Select the Exit icon to exit the menu category and return to the
previous menu.
EDIT
PARAMETER
Select the Edit Parameter icon to select the highlighted parameter
for editing.
PREVIOUS
Select the Previous icon to scroll through the parameters. Once a
parameter is highlighted, edit the parameter by pressing the Edit
Parameter icon.
NEXT
Select the Next icon to scroll through the parameters. Once a
parameter is highlighted, edit the parameter by pressing the Edit
Parameter icon.
Simplified User Interface
By enabling the Simplified User Interface, users are exposed to only the most common
oximeter features, while the all the remaining settings remain available behind password
protection.
FIRST PAGE
NEXT MENU
PAG E
Press the Next Menu Page button to access the second page of
selections.
Press and hold the button for 8 seconds to toggle between
Traditional User Interface and the Simplified User Interface. Enter
password when prompted (refer to Section 4 Password Operation).
The Radical-7 will retain this setting after a power cycle.
ALARM MENU
Press the Alarm Menu Access button to enter the alarm settings
menu.
APOD
MAX
SENSITIVITY
Press the Sensitivity button to toggle between the Normal, APOD
and Maximum Sensitivity modes. Use the Normal Sensitivity set-
ting for typical monitoring purposes. Use the APOD setting where
there is a high probability of the sensor becoming detached. Use
the Maximum Sensitivity setting for patients with low perfusion or
when the low perfusion message is displayed on the screen in
APOD or normal sensitivity mode. The default is APOD.
CAUTION: WHEN USING THE MAXIMUM SENSITIVITY SET-
TING, THE PERFORMANCE OF THE SENSOR OFF DETEC-
TION MAY BE COMPROMISED.
Note: In “Custom” mode the unit will remain in Normal or APOD
setting after a power cycle. Maximum Sensitivity will automatically
reset to Normal Sensitivity after a power cycle.
In “Neo” or “Adult” mode the unit will reset the sensitivity to the
hospital specified setting (Normal or APOD) after a power cycle.
T
REND
GRAPH
TREND GRAPH
Press the Trend Toggle button to alternate between SpO
2
, BPM, PI,
SpMet and SpCO Quick Trend displays
SECOND PAGE
NEXT MENU
PAG E
Press the Next Menu Page button to access the first page of
selections.
Press and hold the button for 8 seconds to toggle between
Traditional User Interface and the Simplified User Interface. Enter
password when prompted. The Radical-7 will retain this setting
after a power cycle.
ALARM MENU
Press the Trend Display button to show the trend data on the dis-
play.
INCREASE
LOUDNESS
Press the Increase Loudness button to increase the volume of the
pulse beep.
Seven levels of volume exist
DECREASE
LOUDNESS
Press the Decrease Loudness button to decrease the volume of
the pulse beep.
operation